COMMODITIES PARAMETERS

 

Voltage

AC 220V/110v  ±10%

Frequency

50/60Hz

Wire plug

GB 3 cores/EN 3 cores/ American

 3 cores

Max Hp

2000VA

Tank Capacity

16L

Class

European Class B standard

Sterilizing temperature

134℃/121℃

Inside Dinebsion

Φ249×450mm

Net weight

53kg

Features

SUN16-II 1800W AC220V/110V±10% 50/60Hz±1Hz 16Litres 445X550X395mm Ø230x360mm 740×550×500mm 47kg

 

Technical parameters

 

Sterilization level

European Class B standard

Sterilizing temperature

134℃/121℃

Special sterilization

Kill active HV,HBV,BSE and spore

Dry procedure

Strong vacuum drying,the residual humidity<0.2%

Display

Widescreen LED 

Test

Bowie&Dick Steam penetration test,  vacuum test , Helix test

Safety

Safety valve, Doble Pressure lock Automatic check.

Sterilizing record

Optional USB and printer (internal or external) to record the sterilization process

Water supply system

3L distiller water tank,3L waste water tank

Clean procedure

Automatic clean inside pipe and steam generator

Instrument containers

5 levels with 3 Round trays 

 

PRODUCT CHARACTER

 

  1.Based on the new class B of international prEN13060, have 3 pulsating vacuum, Vacuum up to -0.9Bar (-0.09Mpa).Suitable to sterilizing wrap, unwrap, solid, porous, hollow device.

  2.Use imported&advancedl6-bit microprocessor,easy to operate. Most suitable for the Dept of Stomatology,Ophthalmology, Surgery and Lab.

  3.With B& D test and vacuum test procedures. To test the penetrability.

  4.Chamber punched by Stainless Steel (#304, thickness: 2.5mm), together with The jet type of steam generator,self-cleaning ensure the efficient sterilizing.

  5.With alarm system for waste water tank, avoid the waste water to enter into the sterilizing circulation, make the sterilizing completely.

  6.Precise LED digital display and advanced self-test system ensure you get the working data freely.

  7.Double lock door system for safety during operating.

  8.with built-in printer .USB optional. The disinfection process data and date automatically print.